xWorking Power Supply | AC220V±10%, 50Hz | Temperature Measurement Range | Room Temperature ~ 370℃ |
---|---|---|---|
Temperature Measurement Accuracy | ≥104℃±2℃ ≤104℃±1℃ | Resolution | 0.1℃ |
Precision | 0. 5% | Heating Speed | In Line With GB/T261-2021 Standard |
Step A | Heating Rate: 5 ~ 6℃/min | Step B | Heating Rate: 1.0 ~ 1.5℃/min |
Step C | Heating Rate:2.5℃~3.5℃/min | Ignition Mode | Electronic Ignition |
Ambient Temperature | (10~40) ℃ | Relative Humidity | ≤ 80% |
Highlight | 500w closed cup flash point tester,automatic closed cup flash point tester,500w flashpoint tester |
SH105BS Automatic closed cup flash point tester
SH105BS Automatic closed cup flash point tester adopts microcomputer technology and large screen LCD LCD display. The instrument warms up according to the standard method, automatic lifting, automatic ignition, automatic display, automatic locking flash point value, automatic printing results. After the test, it can automatically cool down and realize the full automation of the working process. Widely used in electric power, petroleum, chemical industry, commodity inspection, scientific research and other departments. Compliance with ISO-2719, ASTM D93, GB / T261-2021 standards.
Flash point
Under the specified test conditions, the test flame causes the sample vapor to fire and spread the flame to the lowest temperature on the liquid surface to 1013 kPa atmosphere.
Compliance with standard GB / T261-2021; ASTMD93
High Precision Petroleum Products Pensky-Martens Closed Cup Flash Point Tester
Flash Point Apparatus with Closed Cup Testing Method
Pensky-Martens Closed Cup Flash Point Tester
Pensky Martens Closed Cup Flash Point Tester ASTM D93 Standard
Technical performance parameters
● Using color LCD large screen display, full Chinese man-machine dialogue interface, no logo keyboard
● Automatically detect atmospheric pressure and correct the influence of atmospheric pressure on the test and calculate the correction value.
● Opening opening, ignition, detecting and printing data are automatically completed, and the test arm automatically rises and falls.
● Electronic ignition, forced air cooling, no gas source, automatic ignition.
● Printer: It is equipped with a miniature printer
● Air pressure detection device: with the air pressure detection device, automatically correct the atmospheric pressure.
● Automatic ignition of the air source is not required to improve the safety of the test.
● Temperature measurement: Range: room temperature: ~370℃
Accuracy: ≥104℃± 2℃ ≤ 104℃± 1℃
Resolution: 0.1℃; accuracy: 0.5%;
● Base parameter: heating speed: in line with GB / T261-2021 standard
● Ignition mode: with automatic electronic ignition, automatic end point stop function
● It has the function of A, B and C.
● Rated power: 500W
● Overall dimensions: 450 ×380× 350 (mm)
● Instrument weight: 19kg
Working power supply | AC220V±10%, 50Hz |
Temperature measurement |
Range: room temperature ~ 370℃ Accuracy: ≥104℃±2℃ ≤104℃±1℃ Resolution: 0.1℃;Precision: 0. 5% |
Base parameters |
Heating speed: in line with GB/T261-2021 standard; Step A: Heating rate: 5 ~ 6℃/min Step B: Heating rate: 1.0 ~ 1.5℃/min Step C : Heating rate:2.5℃~3.5℃/min Ignition mode: electronic ignition |
Stirrer: Electric stirring device |
(1) 90 ~ 120 RPM, suitable for test step A.C (2) 250±10 RPM, suitable for test step B. |
Ambient temperature: (10~40) ℃ Relative humidity: ≤ 80% | |
Power consumption of the whole machine: | not more than 500W. |
Instrument size: overall size: 450 * 380 * 350 (mm) about 19kg |
Packing list
S/N | Name | Quantity | Unit | remarks |
1 | Host | 1 | set | |
2 | Oil cup | 1 | piece | |
3 | AC power cable | 1 | piece | |
4 | Printer paper | 1 | piece | |
5 | Instruction manual | 1 | piece | |
6 | Packing list | 1 | copy | |
7 | Certificate of qualification | 1 | copy |